Analysis and treatment of common error codes of Noy helium mass spectrometer leak detector
Date:2022-07-11 Views:8946

Error Code |
Meaning |
Cause |
Inspection and Elimination |
10 |
Too many repeated evacuation times during standard leakage testing |
1: The pressure at the leak detection port cannot be pumped down, the leak detection port is not properly connected or there is a fault with the leak hole. 2: Valve V2 is not opened or not fully opened. |
1: Check for large leaks at the leak detection port and test the leak holes. 2: Check Valve V2. |
16 |
Startup timeout |
The molecular pump cannot be powered on normally. |
Check the molecular pump. |
20 |
Anode high voltage error |
The lamp board detects an error in the anode voltage of the ion source. |
Inspect the anode power supply, ion source cable, and filament board. |
21 |
Emission current is low |
The lamp board detects a low emission current of the ion source. |
Press the stop button and restart. |
22 |
Emission current is high |
The lamp board detects a high emission current of the ion source. |
Press the stop button and restart. |
23 |
Lamp board 24V fault |
The 24V voltage of the lamp board is abnormal. |
Inspect the lamp board. |
24 |
80V voltage fault |
The 80V voltage of the lamp board is abnormal. |
Inspect the lamp board. |
26 |
Molecular pump is not ready |
The molecular pump has not been started up completely. |
Check the fore-vacuum and judge according to the molecular pump itself. |