Common faults and solutions of helium mass spectrometer leak detector

Date:2019-09-26 Views:4981

According to the usage feedback of Nuoyi's customers on the helium mass spectrometer leak detector, for several faults with a high failure rate during the usage process and their handling methods, you can refer to the following table:

Category

Fault Phenomenon

Inspection and Analysis Method

Power Supply

The leak detector does not operate after powering on

Is there AC220V at the power socket?

Is the power connection good?

Does the power switch output 220V?

Is the fuse blown?

No display after powering on

Check whether the connection cable of the display module is loose

Mechanical Pump

There is a display after powering on, but the mechanical pump does not rotate

Check whether the mechanical pump switch is turned on;

Check whether the power cable of the mechanical pump is reliably connected

The mechanical pump makes an abnormal noise

Check each connection clamp of the mechanical pump;

Check whether the leak detection port is sealed;

Check whether the fan protective cover of the mechanical pump is deformed;

The mechanical pump is not OK

Check whether there is a large leak;

Check whether the mechanical pump oil is solidified

Ion Source

The filament is not powered

Check whether the molecular pump has been started up completely;

Check whether the ion source socket is loose

The ion source is powered repeatedly

Check the wire sequence of the ion source socket to see if it is inserted wrongly